The difference type between Class N autoclave and Class B autoclave (Vacuum autoclave and Non-Vacuum autoclave)

2016-05-16

The difference type between Class N autoclave and Class B autoclave

 

Briefly speaking, it matters with whether the steam has the ability to sterilize the items’ depth and Aperture.
 

  • The autoclave without vacuum function: It’s mainly utilized in the flash application for raw instruments.
  • The autoclave with vacuum function: It will make the sterilization of wrapped hollow instruments more complete.



In fact, the times of vacuum don’t mean that much. It’s more about whether the sterilization items could pass the trials to prove its sterilizing ability. (With more times of vacuum, we simply hope to approach the demanded parameter more. There are odds to reach the parameter with one time vacuum.)

 

Type N TYPE (Gravity Displacement) B TYPE(Air Displacement:Vacuum)
Pre-Vacuum(Cold air removal)  
Post-Vacuum(Hot air removal)  
Drying effect   BEST
Flash  
Single Package(pouch)  
Single Package(cloth)  
Double Package(cloth)  


According to different instruments’ length and aperture (as beneath), vacuum autoclave needs to reach the standard of EN13060’s Hollow A Test. EQUS makes use of GKE (Germany)’s sterilization detect tube to perform product detection. EQUS vacuum autoclaves (TAT-202B, TAT-203C, TEV-17) all pass European BD-Test and American BD-Test.

 

GKE vacuum test color Vacuum test type
The difference type between Class N autoclave and Class B autoclave

*Hallow A test:Hollow A test described in EN 867-5 required as an additional type test in EN 285.

The difference type between Class N autoclave and Class B autoclave *Europen BD-Test: Air removal and steam penetration test validated according to the test method of EN ISO 11140-1 + 4 with reference to the 7 kg test pack in EN 285
The difference type between Class N autoclave and Class B autoclave

*American BD-Test: Air removal test according to AAMI/ANSI ST79 (4 kg cotton pack) and validated according to the test method of ISO 11140-1 + 5.




*Source by GKE

 

 